Rapid Determination of Acid Phosphatase by a Surface Acoustic Wave Impedance Sensor

A new type of surface acoustic wave impedance sensor based on a conductance-surface acoustic wave frequency reponse was applied to assay the activity of acid phosphatase. Using cytidine-5′-monophosphate (5′-CMP) as a substrate, the Michaelis constant and the corresponding maximum initial rate were e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical letters 1996-01, Vol.29 (1), p.59-72
Hauptverfasser: Wang, R., Cai, Q. Y., Wu, L., Nie, L. H., Yao, S. Z.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new type of surface acoustic wave impedance sensor based on a conductance-surface acoustic wave frequency reponse was applied to assay the activity of acid phosphatase. Using cytidine-5′-monophosphate (5′-CMP) as a substrate, the Michaelis constant and the corresponding maximum initial rate were estimated to be 4.13×10 −3 and 9827 Hz/min, respectively. A linear relationship between frequency response and enzyme concentration was obtained in the range up to 250 μg/ml of acid phosphatase, at 25°C and pH 4.5. The experimental detection limit of acid phosphatase (within 1 min) was 0.9 min/ml and the recovery of the sensor system ranged from 94% to 106% (n = 6). Some other factors such as temperature, pH value and the effects of some metal ions were also discussed.
ISSN:0003-2719
1532-236X