A Passive Quench Protection Design for the 9.4 T MRI Superconducting Magnet
A passive quench protection design of the 9.4 T whole-body magnetic resonance imaging superconducting magnet is proposed. The design of the coil subdivision with shunt resistors is introduced. The selection of the configuration of the heater network is detailed. The optimization of the geometric par...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on applied superconductivity 2014-06, Vol.24 (3), p.1-5 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A passive quench protection design of the 9.4 T whole-body magnetic resonance imaging superconducting magnet is proposed. The design of the coil subdivision with shunt resistors is introduced. The selection of the configuration of the heater network is detailed. The optimization of the geometric parameter of the heater strips and the thickness of the insulation binding outside the heater strips are discussed. A winding sequence that could accelerate the quench propagation in the compensating coil is proposed. The calculation results show that the optimized quench protection design can guarantee the safety of the magnet in the case of any coil as the quench initiation coil. |
---|---|
ISSN: | 1051-8223 1558-2515 |
DOI: | 10.1109/TASC.2013.2283812 |