A New Potentiometric Membrane Sensor Responsive to Uric Acid
A new potentiometric membrane sensor responsive to uric acid has been developed. The membrane sensor is based on aliquat-urate, as an ion pair, and 2-NPOE as a plasticizer. The sensor exhibits fast, stable, and Nernstian response for the hydrogen urate over the concentration range 10 −2 -10 −5 M and...
Gespeichert in:
Veröffentlicht in: | Analytical letters 1997-01, Vol.30 (12), p.2175-2187 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new potentiometric membrane sensor responsive to uric acid has been developed. The membrane sensor is based on aliquat-urate, as an ion pair, and 2-NPOE as a plasticizer. The sensor exhibits fast, stable, and Nernstian response for the hydrogen urate over the concentration range 10
−2
-10
−5
M and pH 5-7. Many substances and ions cause no interference. A standard addition method has been used for determination of uric acid in serum, urine, and commercial standard uric acid and the results of determination were 3.0 to 7.2 mg%, 0.21 to 1.04 mg% and 4.5 × 10
−5
to 4.9 × 10
−3
mg%, respectively. |
---|---|
ISSN: | 0003-2719 1532-236X |
DOI: | 10.1080/00032719708001731 |