Design and calibration of electric field probes in the range 10 - 120 MHz
In view of potential thermal hazards, there is a need to determine the specific absorption rate (SAR) distributions associated with radiofrequency coils used in magnetic resonance imaging (MRI) (typically 10-120 MHz). Electric field (E-field) distributions in tissue-equivalent phantoms may be determ...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 1997-07, Vol.42 (7), p.1387-1394 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In view of potential thermal hazards, there is a need to determine the specific absorption rate (SAR) distributions associated with radiofrequency coils used in magnetic resonance imaging (MRI) (typically 10-120 MHz). Electric field (E-field) distributions in tissue-equivalent phantoms may be determined using a probe comprising a dipole antenna and a detector. The geometry of the dipole dictates the sensitivity of the device, thus two designs are discussed in this paper. Both probes are compact, have a spatial resolution of 2.5 cm3, operate at MR frequencies and have a response independent of the dielectric characteristics of the phantom material. Calibration of these probes requires a system capable of producing a known E-field both in air and in a tissue-like medium at frequencies between 10 and 120 MHz. Transverse electromagnetic wave (TEM) cells answering these specifications are described and the calibration procedure outlined. Accurately calibrated E-field probes can make field measurements in phantoms which can be used to verify predictions from numerical models. These numerical techniques may then be used to predict E-fields, and hence SAR, in patients. |
---|---|
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/0031-9155/42/7/011 |