Dosimetry of Exposure of Patients to Pulsed Gradient Magnetic Fields in MRI
In magnetic resonance imaging, time-varied gradient magnetic fields may stimulate nerves and muscles by inducing electric fields and currents in patients, which may potentially cause health problem. In this study, realistic transverse x -, y -gradient coils and longitudinal z -gradient coil were num...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on magnetics 2011-10, Vol.47 (10), p.3841-3844 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In magnetic resonance imaging, time-varied gradient magnetic fields may stimulate nerves and muscles by inducing electric fields and currents in patients, which may potentially cause health problem. In this study, realistic transverse x -, y -gradient coils and longitudinal z -gradient coil were numerically designed. The magnetic flux density, current density and electric fields in a 3-D real human man model were calculated by employing impedance method. The exposure levels have been compared with the guidelines of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). It was found that the spatially averaged magnetic flux density greatly exceeds the ICNIRP reference level, and the maximum value of the 1-cm 2 -averaged induced current density in central nervous system greatly exceeds the ICNIRP basic restriction. The strongest levels of exposure are observed when all three gradient coils are operated simultaneously. |
---|---|
ISSN: | 0018-9464 1941-0069 |
DOI: | 10.1109/TMAG.2011.2157086 |