Determination of Cyclosporine A by High-Performance Liquid Chromatography with Aryl Oxalate Chemiluminescence Detection
Monitoring of cyclosporine A in 40 whole blood samples from renal transplantation patients by high performance liquid chromatography with aryl oxalate-sulforhodamine 101 chemiluminescence detection is described. These data were compared with fluorescence polarized immunoassay (Cyclosporine-SP Dynapa...
Gespeichert in:
Veröffentlicht in: | Analytical letters 1998-02, Vol.31 (4), p.621-629 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Monitoring of cyclosporine A in 40 whole blood samples from renal transplantation patients by high performance liquid chromatography with aryl oxalate-sulforhodamine 101 chemiluminescence detection is described. These data were compared with fluorescence polarized immunoassay (Cyclosporine-SP Dynapack
TM
using monoclonal antibody for whole blood), with good results (r = 0.88, y = 0.80 + 5.07). These values are the same as reported by high performance liquid chromatography with photometric detection. By using a solid phase extraction cartridge, cyclosporine A was analyzed within 30 min. The analysis time could be shortened to one tenth that previously reported. The proposed method has the same sensitivity as fluorescence polarized immunoassay. |
---|---|
ISSN: | 0003-2719 1532-236X |
DOI: | 10.1080/00032719808001866 |