AMS 14C Sample Preparation at the KCCAMS/UCI Facility: Status Report and Performance of Small Samples

We present an overview of accelerator mass spectrometry (AMS) radiocarbon sample preparation and measurements, describing the technical upgrades that now allow us to routinely obtain 0.2–0.3% precision for 1-mg carbon samples. A precision of ∼1% on samples with 100 μg of carbon can also be achieved....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiocarbon 2007, Vol.49 (2), p.255-269
Hauptverfasser: Santos, G M, Moore, R B, Southon, J R, Griffin, S, Hinger, E, Zhang, D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present an overview of accelerator mass spectrometry (AMS) radiocarbon sample preparation and measurements, describing the technical upgrades that now allow us to routinely obtain 0.2–0.3% precision for 1-mg carbon samples. A precision of ∼1% on samples with 100 μg of carbon can also be achieved. We have also developed graphitization techniques and AMS procedures for ultra-small samples (down to 0.002 mg of carbon). Detailed time series are presented for large and small aliquots of standards such as NIST OX-I and OX-II; FIRI-C and -D; IAEA-C6, -C7 and -C8; and 14C-free samples.
ISSN:0033-8222
1945-5755
DOI:10.1017/S0033822200042181