Critical assessment of fluorescence polarization measurements with a FACS IV cell sorter

The usefulness and limitations of the Becton–Dickinson fluorescence‐activated cell sorter FACS IV for fluorescence polarization measurements were examined. A set of tests to determine the characteristics of the detection geometry, the optical properties of the beam splitter, and the capability to pr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1988-09, Vol.59 (9), p.1928-1932
Hauptverfasser: Muller, Claude P., Krabichler, Gert
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The usefulness and limitations of the Becton–Dickinson fluorescence‐activated cell sorter FACS IV for fluorescence polarization measurements were examined. A set of tests to determine the characteristics of the detection geometry, the optical properties of the beam splitter, and the capability to process fluorescence polarization data is presented. Recommendations are provided for correcting instrumental deficiencies.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1140053