Inter-Element Interference Studies in the Determination of Trace Amounts of Plutonium by Inductively Coupled Plasma - Atomic Emission Spectrometry
Trace amounts of plutonium in aqueous solutions have been determined by an inductively coupled argon plasma-atomic emission spectrometric (ICP-AES) technique. Extensive studies have been carried out on interelement interference effects for two plutonium lines, viz., 300.06 nm and 453.6 nm, due to th...
Gespeichert in:
Veröffentlicht in: | Analytical letters 1990-11, Vol.23 (11), p.2095-2106 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Trace amounts of plutonium in aqueous solutions have been determined by an inductively coupled argon plasma-atomic emission spectrometric (ICP-AES) technique. Extensive studies have been carried out on interelement interference effects for two plutonium lines, viz., 300.06 nm and 453.6 nm, due to the presence of uranium, thorium, lanthanides and other metallic elements in the sample solutions. The analytical range obtained in the present case is 0.1-100 μg/ml with a precision of better than 1% R.S.D. The method ie checked for accuracy by analyzing two wellcharacterizd reference rtandards. |
---|---|
ISSN: | 0003-2719 1532-236X |
DOI: | 10.1080/00032719008052552 |