Characterization of changes in therapeutic ultrasound transducer performance over time using the angular spectrum method
Strongly focused large aperture transducers used in high-intensity focused ultrasound treatments are prone to manufacturing defects and degradation. Current methods for evaluating transducer quality measure only bulk physical changes of transducers. We have determined the pressure distribution at th...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 2007-05, Vol.54 (5), p.1028-1035 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Strongly focused large aperture transducers used in high-intensity focused ultrasound treatments are prone to manufacturing defects and degradation. Current methods for evaluating transducer quality measure only bulk physical changes of transducers. We have determined the pressure distribution at the transducer surface, using the angular spectrum method, to detect defects of the transducer. Three therapeutic transducers were investigated. The pressure distribution at the focal plane of each transducer was measured and input into a back-projection algorithm to calculate the pressure distribution at the transducer surface. A number of scan window sizes were used for the pressure distribution measurement at the focal plane to determine the effect on the resolution of the calculated pressure distribution at the transducer surface. Results showed that one transducer might have suffered manufacturing defects. The second transducer degraded over 1 year of use with one half of the transducer suffering a partial loss of efficiency. The third transducer remained unchanged over 1 year. The scan window of 40 mm times 40 mm at the focal plane was required to identify defects 6 mm in diameter on the transducer surface. The results demonstrate that the angular spectrum method could be a useful tool for evaluating transducer quality |
---|---|
ISSN: | 0885-3010 1525-8955 |
DOI: | 10.1109/TUFFC.2007.348 |