A high dynamic range current-mode amplifier for computed tomography
In this paper, we show a high dynamic range current-mode detector for computed tomography application. A regulated current mirror structure has been implemented at pixel level that provides with 17 bits dynamic range and a noise floor below 3 pARMS. Nonlinearity is kept below 2% and signal bandwidth...
Gespeichert in:
Veröffentlicht in: | IEEE journal of solid-state circuits 2006-07, Vol.41 (7), p.1615-1619 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this paper, we show a high dynamic range current-mode detector for computed tomography application. A regulated current mirror structure has been implemented at pixel level that provides with 17 bits dynamic range and a noise floor below 3 pARMS. Nonlinearity is kept below 2% and signal bandwidth is higher than 10 kHz. A test structure with 4times4 pixel array is presented is this paper. Both photodiode and current mode amplifier have been integrated into the same CMOS standard process |
---|---|
ISSN: | 0018-9200 1558-173X |
DOI: | 10.1109/JSSC.2006.873679 |