A miniaturized lock-in amplifier design suitable for impedance measurements in cells [biological cells]

A comparison of the performance for impedance analysis of biological cells between a newly fabricated miniature lock-in amplifier and an industry standard Stanford Research SR830 lock-in amplifier is described. Experimental results compare measurement error over a range of characteristic model imped...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Moe, A.E., Marx, S.R., Bhinderwala, I., Wilson, D.M.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A comparison of the performance for impedance analysis of biological cells between a newly fabricated miniature lock-in amplifier and an industry standard Stanford Research SR830 lock-in amplifier is described. Experimental results compare measurement error over a range of characteristic model impedances from 1 kHz to 10 MHz. Within the functional frequency range of the SR830, the miniaturized lock-in amplifier performs with approximately 30% less precision than the SR830. At higher frequencies, that are more suitable for impedance analysis in biological cells, however, the miniaturized lock-in continues to perform with less than 10% average error while the SR830 fails to operate entirely. With consistent performance at higher frequencies, this miniaturized lock-in amplifier is a desirable instrument for use in biological impedance analysis.
DOI:10.1109/ICSENS.2004.1426139