Microfracture of a Baclofen Pump Catheter with Intermittent Under- and Overdose

An 8-year-old boy with an implanted intrathecal baclofen pump presented with intermittent signs of baclofen under- and overdose. Pump interrogation, pump volumes, plain X-rays, catheter contrast fluoroscopy and CT of the intrathecal catheter tip were unremarkable. A microfracture of the extradural c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pediatric neurosurgery 2003-09, Vol.39 (3), p.144-148
Hauptverfasser: Dawes, William J., Drake, James M., Fehlings, Darcy
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An 8-year-old boy with an implanted intrathecal baclofen pump presented with intermittent signs of baclofen under- and overdose. Pump interrogation, pump volumes, plain X-rays, catheter contrast fluoroscopy and CT of the intrathecal catheter tip were unremarkable. A microfracture of the extradural catheter was only seen at a magnification of ×20 and confirmed by scanning electron microscopy. Symptoms resolved following replacement. Microfracture of the intrathecal catheter of baclofen pump systems may present puzzling symptomatology and be difficult to diagnose by conventional means.
ISSN:1016-2291
1423-0305