Measurements of MTF and SNR( f ) using a subtraction method in MRI
A method was developed for accurate measurement of the modulation transfer function (MTF) and signal-to-noise ratio in the spatial frequency domain (SNR(f)) of magnetic resonance images (MRI). The MTF was calculated from the complex images of a line object which were obtained by the subtraction of t...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 2002-08, Vol.47 (16), p.2961-2972, Article 311 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A method was developed for accurate measurement of the modulation transfer function (MTF) and signal-to-noise ratio in the spatial frequency domain (SNR(f)) of magnetic resonance images (MRI). The MTF was calculated from the complex images of a line object which were obtained by the subtraction of two separately acquired data sets of a specially designed phantom with a sliding sheet. Moreover, the SNR(f) was calculated from the MTF and Wiener spectrum, both of which were determined using the same phantom configuration. The MTFs and SNR(f)s in the conventional spin-echo (SE) and turbo SE, in which the effective echo time was set to the first echo, were evaluated by changing the T2 of the phantom and the echo train length. The MTFs in the positive and negative frequencies indicated the effect of the k-space trajectory for each pulse sequence. SNR(f)s gave spatial frequency information that was not obtained with conventional methods. In this method, the influence of image nonuniformity and unwanted artefacts (edge and ghost) could be eliminated. An analysis of the MTF and the SNR in the spatial frequency domain provides additional information for the assessment of image quality in MRI. |
---|---|
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/0031-9155/47/16/311 |