An adiabatic calorimeter for small samples: The solid–liquid system naphthalene–durene

The construction and calibration of an adiabatic calorimeter for samples of about 0.4g is described. The calorimeter works between 10 and 420K, the accuracy is 0.5% for heat capacity measurements and 0.2% for latent energy effect like the enthalpy of fusion. The solid–liquid phase diagram of the sys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Thermochimica acta 2002-02, Vol.383 (1-2), p.13-19
Hauptverfasser: van Miltenburg, J.C., van den Berg, G.J.K., van Genderen, A.C.G.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The construction and calibration of an adiabatic calorimeter for samples of about 0.4g is described. The calorimeter works between 10 and 420K, the accuracy is 0.5% for heat capacity measurements and 0.2% for latent energy effect like the enthalpy of fusion. The solid–liquid phase diagram of the system naphthalene–durene is presented. The phase diagram has been determined by three measurements, the interrupted way of measuring in adiabatic calorimetry was used to calculate the liquid composition at different enthalpy levels. The phase diagram is an example of an almost ideal eutectic system. The excess Gibbs energy of the liquid phase at 360K was found to be in the order of 100Jmol−1 for the equimolar mixture.
ISSN:0040-6031
1872-762X
DOI:10.1016/S0040-6031(01)00701-8