Phase-Doppler Diagnostic Studies of Primary and Tertiary Aerosols Produced by a High-Efficiency Nebulizer

Two phase-Doppler instruments are used to diagnose the primary and tertiary aerosols prior to injection into a high-temperature plasma. Measured parameters include droplet size and velocity distributions, size-velocity correlation, span of aerosols, droplet number density, volume flux and percent vo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 1994-10, Vol.66 (19), p.3233-3242
Hauptverfasser: Liu, Huiying, Montaser, Akbar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two phase-Doppler instruments are used to diagnose the primary and tertiary aerosols prior to injection into a high-temperature plasma. Measured parameters include droplet size and velocity distributions, size-velocity correlation, span of aerosols, droplet number density, volume flux and percent volume of aerosol droplets under eight mu m.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac00091a037