Metalloprotein separation and analysis by directly coupled size exclusion high-performance liquid chromatography inductively coupled plasma mass spectroscopy
The feasibility of using directly coupled size exclusion high-performance liquid chromatography inductively coupled plasma mass spectroscopy ( HPLC ICP-MS ) for the separation and subsequent elemental analysis of metalloproteins in biological samples has been studied. Data, on up to eight elements,...
Gespeichert in:
Veröffentlicht in: | Analytical biochemistry 1990-05, Vol.186 (2), p.187-201 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The feasibility of using directly coupled size exclusion high-performance liquid chromatography inductively coupled plasma mass spectroscopy (
HPLC
ICP-MS
) for the separation and subsequent elemental analysis of metalloproteins in biological samples has been studied. Data, on up to eight elements, was acquired simultaneously and the reconstructed elemental profiles from the chromatographed samples were quantified by flow injection analysis. Absolute and relative detection limits, reproducibility, operational dynamic range, and linearity of response were initially evaluated by analyzing standards of metallothionein protein of known elemental composition for Cd, Zn, and Cu. There was evidence of displacement of Zn from the protein during chromatography and the substitution of Cu sequestered from the mobile phase. Cd associated with the protein was fully recovered during chromatography. Memory effects, due to protein adsorption to the glassware in the torch box, were minimal and there was no degradation of the resolution of the chromatographed peak during extended transport through the
HPLC
ICP-MS
interface. The versatility of the technique has been demonstrated by the quantitative multielement analysis of cytosolic metal-binding proteins separated from the polychaete worm
Neanthes arenaceodentata. Fidelity of analysis has been demonstrated by two independent procedures: first, by comparing the elemental profiles obtained by directly aspirating the HPLC eluant into the ICP-MS to those obtained by collecting fractions and quantifying the metal content of the proteins in the conventional analytical mode; second, by comparing the stable isotopic profiles for
114Cd obtained by simultaneous ICP-MS analysis with radiometric profiles of
109Cd obtained by counting radioactivity associated with collected fractions. The virtues and limitations of ICP-MS as a multielement HPLC detector for the analysis of metalloproteins are described and compared to more established techniques using other element specific or multielement detectors. |
---|---|
ISSN: | 0003-2697 1096-0309 |
DOI: | 10.1016/0003-2697(90)90066-I |