Calibration of a monochromator/spectrometer system for the measurement of photoelectron angular distributions and branching ratios

We describe the techniques used in calibrating a monochromator/spectrometer system for gas-phase photoelectron angular distribution and branching ratio measurements. We report a self-consistent set of values for the Ne 2p, Ar 3p, Kr 4 p 3 2 and 4 p 1 2 , and Xe 5 p 3 2 and 5 p 1 2 photoelectron asym...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 1986-05, Vol.246 (1), p.782-786
Hauptverfasser: Southworth, S.H., Parr, A.C., Hardis, J.E., Dehmer, J.L., Holland, D.M.P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe the techniques used in calibrating a monochromator/spectrometer system for gas-phase photoelectron angular distribution and branching ratio measurements. We report a self-consistent set of values for the Ne 2p, Ar 3p, Kr 4 p 3 2 and 4 p 1 2 , and Xe 5 p 3 2 and 5 p 1 2 photoelectron asymmetry parameters and for the Kr 4 p 3 2 : 4 p 1 2 and Xe 5 p 3 2 : 5 p 1 2 branching ratios for the kinetic energy regions from threshold to approximately 15 eV.
ISSN:0168-9002
1872-9576