Spot-based measurement of the brighter-fatter effect on a Roman Space Telescope H4RG detector and comparison with flat-field data
We present the measurement and characterization of the brighter-fatter effect (BFE) on a NASA Roman Space Telescope development Teledyne H4RG-10 near-infrared detector using laboratory measurements with projected point sources. After correcting for other interpixel non-linearity effects such as clas...
Gespeichert in:
Veröffentlicht in: | Journal of instrumentation 2024-03, Vol.19 (3), p.P03015 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present the measurement and characterization of the
brighter-fatter effect (BFE) on a NASA Roman Space Telescope
development Teledyne H4RG-10 near-infrared detector using laboratory
measurements with projected point sources. After correcting for
other interpixel non-linearity effects such as classical
non-linearity and inter-pixel capacitance, we quantify the magnitude
of the BFE by calculating the fractional area change per electron of
charge contrast. We also introduce a mathematical framework to
compare our results with the BFE measured on similar devices using
autocorrelations from flat-field images. We find an agreement of
18 ± 5% between the two methods. We identify potential sources
of discrepancy and discuss future investigations to characterize and
address them. |
---|---|
ISSN: | 1748-0221 1748-0221 |
DOI: | 10.1088/1748-0221/19/03/P03015 |