Metal Artifact Reduction for Orthopedic Prosthesis in Lower Extremity CT Venography: Evaluation of Image Quality and Vessel Conspicuity

Purpose This study evaluates the image quality of lower extremity CT venography reconstructed with orthopedic metal artifact reduction (O-MAR) in patients with unilateral or bilateral metallic prostheses in the hip or knee. Methods This retrospective study was approved by our institutional review bo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular and interventional radiology 2019-11, Vol.42 (11), p.1619-1626
Hauptverfasser: Woo, Hyunsik, Chai, Jee Won, Choi, Young Ho, Jin, Kwang Nam
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose This study evaluates the image quality of lower extremity CT venography reconstructed with orthopedic metal artifact reduction (O-MAR) in patients with unilateral or bilateral metallic prostheses in the hip or knee. Methods This retrospective study was approved by our institutional review board, and informed consent was waived. Twenty-nine patients of lower extremity CT with 51 metallic hip or knee prostheses were reconstructed to both standard CT images and O-MAR images. The subjective image quality and vessel conspicuity for both images were evaluated by two readers using five-point scales (0–4). Vessel conspicuity scores of 3 or 4 were considered diagnostically acceptable. Image noise was measured in the air and subcutaneous fat. Results O-MAR images showed significantly higher scores of subjective image quality ( p  
ISSN:0174-1551
1432-086X