A Study of the Residual Gas Composition in the Vacuum System of the Cyclotron of the Ioffe Physical Technical Institute

Mass spectra of residual gas molecules in vacuum system of the cyclotron of the Ioffe Physical Technical Institute (St. Petersburg) were obtained under various experimental conditions using a compact FT-200 time-of-flight mass spectrometer designed and manufactured at the institute. The measurements...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Technical physics letters 2019-08, Vol.45 (8), p.827-830
Hauptverfasser: Aruev, N. N., Kozlovskii, M. A., Kudoyarov, M. F., Patrova, M. Ya, Romanov, P. A., Tyukal’tsev, R. V., Fedichkin, I. L., Filippov, S. V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Mass spectra of residual gas molecules in vacuum system of the cyclotron of the Ioffe Physical Technical Institute (St. Petersburg) were obtained under various experimental conditions using a compact FT-200 time-of-flight mass spectrometer designed and manufactured at the institute. The measurements were performed in the vacuum chamber of cyclotron and in a terminal part of the ion guiding channel near the sample irradiation chamber. It is established that main contributions to residual gas composition in the cyclotron vacuum chamber are related to molecules of the vacuum oil (up to 82 vol %) and water molecules (up to 15–17 vol %), Water molecules also account for the main contribution (up to 63 vol %) to residual gas composition in the terminal part of the ion guiding channel near the sample irradiation chamber with a polymer film. The obtained data provide a basis for upgrade of the cyclotron complex of the Ioffe Physical Technical Institute.
ISSN:1063-7850
1090-6533