Evaluation of deformed image-based dose calculations for adaptive radiotherapy of nasopharyngeal carcinoma
The ultimate goal of adaptive radiotherapy (ART) is to deliver truly customized radiation treatments. Currently, the quality of cone-beam computed tomography (CBCT) images is still inferior to that of conventional CT images in contour delineations and dose calculations for replanning purposes. This...
Gespeichert in:
Veröffentlicht in: | Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2017, Vol.42 (4), p.273-281 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The ultimate goal of adaptive radiotherapy (ART) is to deliver truly customized radiation treatments. Currently, the quality of cone-beam computed tomography (CBCT) images is still inferior to that of conventional CT images in contour delineations and dose calculations for replanning purposes. This retrospective study aims to evaluate the dosimetric accuracy of using deformed conventional CT images for dose calculations, in the hope of inferring the feasibility of ART using planning CT (PCT) images that deformed to up-to-date CBCT images for patients with nasopharyngeal carcinoma (NPC). Thirty consecutive patients with NPC who had undergone 1 replan in their radiotherapy treatments were selected. The pretreatment PCT images were deformed to match the mid-treatment PCT images by deformable image registration. The same volumetric modulated arc therapy plan was then calculated on the deformed PCT images. The resulting dose distributions and dose volume histograms of the tumors and organs at risk (OARs) were compared with the original plan. Five dose levels, D98%, D95%, D50%, D5%, and D2%, were recorded for 9 NPC targets. Four dose levels, Dmax, D10%, D50%, and Dmean, were recorded for 15 OARs. The greatest percentage difference in observed dose for D98%, D95%, D50%, D5%, and D2% of the targets were 1.71%, 1.55%, 0.64%, 0.97%, and 1.13%, respectively. The greatest percentage difference in observed dose for Dmax, D10%, D50%, and Dmean of the OARs were -26.51% (left optic nerve), -17.06% (left optic nerve), 56.70% (spinal cord), and 18.97% (spinal cord), respectively. In addition, 29 of 45 (64%) dosimetric end points of the targets showed statistically significant dose differences (p |
---|---|
ISSN: | 0958-3947 1873-4022 |
DOI: | 10.1016/j.meddos.2017.06.001 |