Spectroscopic Measurements of L X-rays with a TES Microcalorimeter for a Non-destructive Assay of Transuranium Elements
Spectroscopic measurement of the L X-rays emitted from transuranium elements is one of the most useful techniques for the non-destructive assays of nuclear materials. In this study, we fabricated a transition-edge-sensor (TES) microcalorimeter using a 5-μm-thick Au absorber and tested its ability to...
Gespeichert in:
Veröffentlicht in: | Journal of low temperature physics 2018-11, Vol.193 (3-4), p.314-320 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Spectroscopic measurement of the L X-rays emitted from transuranium elements is one of the most useful techniques for the non-destructive assays of nuclear materials. In this study, we fabricated a transition-edge-sensor (TES) microcalorimeter using a 5-μm-thick Au absorber and tested its ability to measure the L X-rays emitted from two transuranium elements, Np-237 and Cm-244 sources. The microcalorimeter was found to successfully measure the L X-rays with an energy resolution (full width at half maximum) below 70 eV. These results confirm that L X-rays can be identified using the proposed TES microcalorimeter to enable non-destructive assays of transuranium elements. |
---|---|
ISSN: | 0022-2291 1573-7357 |
DOI: | 10.1007/s10909-018-1953-9 |