Subsecond and Submillimeter Resolution Positional Verification for Stereotactic Irradiation of Spinal Lesions
Purpose Spine stereotactic body radiation therapy (SBRT) requires highly accurate positioning. We report our experience with markerless template matching and triangulation of kilovoltage images routinely acquired during spine SBRT, to determine spine position. Methods and Materials Kilovoltage image...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 2016-04, Vol.94 (5), p.1154-1162 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose Spine stereotactic body radiation therapy (SBRT) requires highly accurate positioning. We report our experience with markerless template matching and triangulation of kilovoltage images routinely acquired during spine SBRT, to determine spine position. Methods and Materials Kilovoltage images, continuously acquired at 7, 11 or 15 frames/s during volumetric modulated spine SBRT of 18 patients, consisting of 93 fluoroscopy datasets (1 dataset/arc), were analyzed off-line. Four patients were immobilized in a head/neck mask, 14 had no immobilization. Two-dimensional (2D) templates were created for each gantry angle from planning computed tomography data and registered to prefiltered kilovoltage images to determine 2D shifts between actual and planned spine position. Registrations were considered valid if the normalized cross correlation score was ≥0.15. Multiple registrations were triangulated to determine 3D position. For each spine position dataset, average positional offset and standard deviation were calculated. To verify the accuracy and precision of the technique, mean positional offset and standard deviation for twenty stationary phantom datasets with different baseline shifts were measured. Results For the phantom, average standard deviations were 0.18 mm for left-right (LR), 0.17 mm for superior-inferior (SI), and 0.23 mm for the anterior-posterior (AP) direction. Maximum difference in average detected and applied shift was 0.09 mm. For the 93 clinical datasets, the percentage of valid matched frames was, on average, 90.7% (range: 49.9-96.1%) per dataset. Average standard deviations for all datasets were 0.28, 0.19, and 0.28 mm for LR, SI, and AP, respectively. Spine position offsets were, on average, −0.05 (range: −1.58 to 2.18), −0.04 (range: −3.56 to 0.82), and −0.03 mm (range: −1.16 to 1.51), respectively. Average positional deviation was |
---|---|
ISSN: | 0360-3016 1879-355X |
DOI: | 10.1016/j.ijrobp.2016.01.006 |