Calibration of time of flight detectors using laser-driven neutron source

Calibration of three scintillators (EJ232Q, BC422Q, and EJ410) in a time-of-flight arrangement using a laser drive-neutron source is presented. The three plastic scintillator detectors were calibrated with gamma insensitive bubble detector spectrometers, which were absolutely calibrated over a wide...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2015-07, Vol.86 (7), p.073308-073308
Hauptverfasser: Mirfayzi, S R, Kar, S, Ahmed, H, Krygier, A G, Green, A, Alejo, A, Clarke, R, Freeman, R R, Fuchs, J, Jung, D, Kleinschmidt, A, Morrison, J T, Najmudin, Z, Nakamura, H, Norreys, P, Oliver, M, Roth, M, Vassura, L, Zepf, M, Borghesi, M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Calibration of three scintillators (EJ232Q, BC422Q, and EJ410) in a time-of-flight arrangement using a laser drive-neutron source is presented. The three plastic scintillator detectors were calibrated with gamma insensitive bubble detector spectrometers, which were absolutely calibrated over a wide range of neutron energies ranging from sub-MeV to 20 MeV. A typical set of data obtained simultaneously by the detectors is shown, measuring the neutron spectrum emitted from a petawatt laser irradiated thin foil.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.4923088