Note: O-ring stack system for electron gun alignment
We present a reliable method for aligning an electron gun which consists of an electron source and lenses by controlling a stack of rubber O-rings in a vacuum condition. The beam direction angle is precisely tilted along two axes by adjusting the height difference of a stack of O-rings. In addition,...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2015-01, Vol.86 (1), p.016110-016110 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present a reliable method for aligning an electron gun which consists of an electron source and lenses by controlling a stack of rubber O-rings in a vacuum condition. The beam direction angle is precisely tilted along two axes by adjusting the height difference of a stack of O-rings. In addition, the source position is shifted in each of three orthogonal directions. We show that the tilting angle and linear shift along the x and y axes as obtained from ten stacked O-rings are ±2.55° and ±2 mm, respectively. This study can easily be adapted to charged particle gun alignment and adjustments of the flange position in a vacuum, ensuring that its results can be useful with regard to electrical insulation between flanges with slight modifications. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4905575 |