Wide dynamic range neutron flux monitor having fast time response for the Large Helical Device
A fast time response, wide dynamic range neutron flux monitor has been developed toward the LHD deuterium operation by using leading-edge signal processing technologies providing maximum counting rate up to ∼5 × 10(9) counts/s. Because a maximum total neutron emission rate over 1 × 10(16) n/s is pre...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2014-11, Vol.85 (11), p.11E114-11E114 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A fast time response, wide dynamic range neutron flux monitor has been developed toward the LHD deuterium operation by using leading-edge signal processing technologies providing maximum counting rate up to ∼5 × 10(9) counts/s. Because a maximum total neutron emission rate over 1 × 10(16) n/s is predicted in neutral beam-heated LHD plasmas, fast response and wide dynamic range capabilities of the system are essential. Preliminary tests have demonstrated successful performance as a wide dynamic range monitor along the design. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.4891049 |