CT-Guided Interventions Using a Free-Hand, Optical Tracking System: Initial Clinical Experience

Purpose The present study was designed to evaluate the geometrical accuracy and clinical applicability of a new, free-hand, CT-guided, optical navigation system. Methods Fifteen procedures in 14 consecutive patients were retrospectively analyzed. The navigation system was applied for interventional...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular and interventional radiology 2013-08, Vol.36 (4), p.1055-1062
Hauptverfasser: Schubert, Tilman, Jacob, Augustinus L., Pansini, Michele, Liu, David, Gutzeit, Andreas, Kos, Sebastian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose The present study was designed to evaluate the geometrical accuracy and clinical applicability of a new, free-hand, CT-guided, optical navigation system. Methods Fifteen procedures in 14 consecutive patients were retrospectively analyzed. The navigation system was applied for interventional procedures on small target lesions, in cases with long needle paths, narrow access windows, or when an out-of-plane access was expected. Mean lesion volume was 27.9 ml, and mean distance to target measured was 107.5 mm. Eleven of 15 needle trajectories were planned as out-of-plane approaches regarding the axial CT plane. Results Ninety-one percent of the biopsies were diagnostic. All therapeutic interventions were technically successful. Targeting precision was high with a mean distance of the needle tip from planned target of 1.98 mm. Mean intervention time was 1:12 h. A statistically significant correlation between angular needle deviation and intervention time ( p  = 0.007), respiratory movement of the target ( p  = 0.008), and body mass index ( p  = 0.02) was detected. None of the evaluated parameters correlated significantly with the distance from the needle tip to the planned target. Conclusions The application of a navigation system for complex CT-guided procedures provided safe and effective targeting within a reasonable intervention time in our series.
ISSN:0174-1551
1432-086X
DOI:10.1007/s00270-012-0527-5