Two-port microwave calibration at millikelvin temperatures
In this work we introduce a system for 2-port microwave calibration at millikelvin temperatures operating at the coldest stage of a dilution refrigerator by use of an adapted thru-reflect-line algorithm. We show that this can be an effective tool for characterizing common 50 Ω microwave components w...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2013-03, Vol.84 (3), p.034704-034704 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this work we introduce a system for 2-port microwave calibration at millikelvin temperatures operating at the coldest stage of a dilution refrigerator by use of an adapted thru-reflect-line algorithm. We show that this can be an effective tool for characterizing common 50 Ω microwave components with better than 0.1 dB accuracy at temperatures that are relevant to many current experiments in superconducting quantum information. |
---|---|