Nanoradian angular stabilization of x-ray optical components

An x-ray free-electron laser oscillator (XFELO) has been recently proposed [ K. Kim , Phys. Rev. Lett. 100 , 244802 ( 2008 ) ]. Angular orientation and position in space of Bragg mirrors of the XFELO optical cavity must be continuously adjusted to compensate for the instabilities and maximize the ou...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2010-05, Vol.81 (5), p.055108-055108-7
Hauptverfasser: Stoupin, Stanislav, Lenkszus, Frank, Laird, Robert, Goetze, Kurt, Kim, Kwang-Je, Shvyd'ko, Yuri
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An x-ray free-electron laser oscillator (XFELO) has been recently proposed [ K. Kim , Phys. Rev. Lett. 100 , 244802 ( 2008 ) ]. Angular orientation and position in space of Bragg mirrors of the XFELO optical cavity must be continuously adjusted to compensate for the instabilities and maximize the output intensity. An angular stability of about 10 nrad (rms) is required [ K. Kim and Y. Shvyd'ko , Phys. Rev. ST Accel. Beams 12 , 030703 ( 2009 ) ]. To approach this goal, a feedback loop based on a null-detection principle was designed and used for stabilization of a high-energy-resolution x-ray monochromator ( Δ E / E ≃ 4 × 10 − 8 , E = 23.7   keV ) and a high-heat-load monochromator. Angular stability of about 13 nrad (rms) has been demonstrated for x-ray optical elements of the monochromators.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.3428722