Evaluation of a 3D Diode Array Dosimeter for Helical Tomotherapy Delivery QA

Abstract The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2010, Vol.35 (4), p.324-329
Hauptverfasser: Feygelman, Vladimir, Ph.D, Opp, D., Ph.D, Javedan, K., M.S, Saini, A.J., Ph.D, Zhang, G., Ph.D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (
ISSN:0958-3947
1873-4022