A Study of Experimental Measurements of Dosimetric Parameters in HDR IR-192 Source
Abstract Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have...
Gespeichert in:
Veröffentlicht in: | Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2010, Vol.35 (4), p.250-254 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them. |
---|---|