Retrievable Vena Cava Filters in Major Trauma Patients: Prevalence of Thrombus Within the Filter

The purpose of this study was to report the prevalence of thrombus within a retrievable vena cava filter inserted prophylactically in major trauma patients referred for filter extraction. Between November 2002 and August 2005, 80 retrievable inferior vena cava filters (68 Optease and 12 Gunther-Tuli...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular and interventional radiology 2008-07, Vol.31 (4), p.785-789
Hauptverfasser: Mahrer, Arie, Zippel, Douglas, Garniek, Alexander, Golan, Gil, Bensaid, Paul, Simon, Daniel, Rimon, Uri
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to report the prevalence of thrombus within a retrievable vena cava filter inserted prophylactically in major trauma patients referred for filter extraction. Between November 2002 and August 2005, 80 retrievable inferior vena cava filters (68 Optease and 12 Gunther-Tulip) were inserted into critically injured trauma patients (mean injury severity score 33.5). The filters were inserted within 1 to 6 (mean 2) days of injury. Thirty-seven patients were referred for filter removal (32 with Optease and 5 with Gunther-Tulip). The indwelling time was 7 to 22 (mean 13) days. All patients underwent inferior vena cavography prior to filter removal. There were no insertion-related complications and all filters were successfully deployed. Forty-three (54%) of the 80 patients were not referred for filter removal, as these patients continued to have contraindications to anticoagulation. Thirty-seven patients (46%) were referred for filter removal. In eight of them (22%) a large thrombus was seen within the filters and they were left in place, all with the Optease device. The other 29 filters (36%) were removed uneventfully. We conclude that the relatively high prevalence of intrafilter thrombi with the Optease filter may be explained by either spontaneous thrombus formation or captured emboli.
ISSN:0174-1551
1432-086X
DOI:10.1007/s00270-008-9294-8