Testing Percutaneous Arterial Closure Devices: An Animal Model
The ovine superficial femoral artery was used for testing the efficacy of percutaneous arterial closure devices (PACDs) in their developmental stage. Two topical devices containing chitostan, one staple-mediated PACD and a porcine small intestinal submucosa plug, were tested by follow-up angiography...
Gespeichert in:
Veröffentlicht in: | Cardiovascular and interventional radiology 2009-03, Vol.32 (2), p.313-316 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The ovine superficial femoral artery was used for testing the efficacy of percutaneous arterial closure devices (PACDs) in their developmental stage. Two topical devices containing chitostan, one staple-mediated PACD and a porcine small intestinal submucosa plug, were tested by follow-up angiography in 37 sheep. Absence or presence of bleeding and time to bleeding cessation were the main criteria for evaluation of PAVD efficacy. The results of these tests directed modification of individual PACDs and improved their efficacy. |
---|---|
ISSN: | 0174-1551 1432-086X |
DOI: | 10.1007/s00270-008-9426-1 |