Use of a synchronization card for XMCD measurements at the XMaS Beamline
X-ray Magnetic Circular Dichroism (XMCD) spectra from a 5 {mu}m polycrystalline Gd foil were measured in transmission across the Gd L{sub 2,3} edges using the Multipurpose Unit for Synchronisation, Sequencing and Triggering (MUSST) card developed at the European Synchrotron Radiation Facility (ESRF)...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | X-ray Magnetic Circular Dichroism (XMCD) spectra from a 5 {mu}m polycrystalline Gd foil were measured in transmission across the Gd L{sub 2,3} edges using the Multipurpose Unit for Synchronisation, Sequencing and Triggering (MUSST) card developed at the European Synchrotron Radiation Facility (ESRF). The circularly polarized x-rays were produced by a diamond phase-plate and their helicity was rapidly reversed through the use of a piezo driven device. The signal measured for each helicity was recorded by the MUSST card synchronized to the helicity reversal. The results are compared to those obtained using an analogue lock-in technique. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.3463352 |