GEANT4 Application for the Simulation of the Head of a Siemens Primus Linac

The Monte Carlo simulation of the head of a Siemens Primus Linac used at Virgen Macarena Hospital (Sevilla, Spain) has been performed using the code GEANT4, version 9.2. In this work, the main features of the application built by our group are presented. They are mainly focused in the optimization o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:AIP conference proceedings 2010, Vol.1231 (1)
Hauptverfasser: Cortes-Giraldo, M. A., Quesada, J. M., Gallardo, M. I.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Monte Carlo simulation of the head of a Siemens Primus Linac used at Virgen Macarena Hospital (Sevilla, Spain) has been performed using the code GEANT4, version 9.2. In this work, the main features of the application built by our group are presented. They are mainly focused in the optimization of the performance of the simulation. The geometry, including the water phantom, has been entirely wrapped by a shielding volume which discards all the particles escaping far away through its walls. With this, a factor of four in the time spent by the simulation can be saved. An interface to read and write phase-space files in IAEA format has been also developed to save CPU time in our simulations. Finally, some calculations of the dose absorption in the water phantom have been done and compared with the results given by EGSnrc and with experimental data obtained for the calibration of the machine.
ISSN:0094-243X
1551-7616
DOI:10.1063/1.3428928