Utility of Megavoltage Fan-Beam CT for Treatment Planning in a Head-And-Neck Cancer Patient with Extensive Dental Fillings Undergoing Helical Tomotherapy

Abstract The purpose of this study was to demonstrate the potential utility of megavoltage fan-beam computed tomography (MV-FBCT) for treatment planning in a patient undergoing helical tomotherapy for nasopharyngeal carcinoma in the presence of extensive dental artifact. A 28-year-old female with lo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2010, Vol.35 (2), p.108-114
Hauptverfasser: Yang, Claus, Ph.D, Liu, Tianxiao, Ph.D, Jennelle, Richard L., M.D, Ryu, Janice K., M.D, Vijayakumar, Srinivasan, M.D, Purdy, James A., Ph.D, Chen, Allen M., M.D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The purpose of this study was to demonstrate the potential utility of megavoltage fan-beam computed tomography (MV-FBCT) for treatment planning in a patient undergoing helical tomotherapy for nasopharyngeal carcinoma in the presence of extensive dental artifact. A 28-year-old female with locally advanced nasopharyngeal carcinoma presented for radiation therapy. Due to the extensiveness of the dental artifact present in the oral cavity kV-CT scan acquired at simulation, which made treatment planning impossible on tomotherapy planning system, MV-FBCT imaging was obtained using the HI-ART tomotherapy treatment machine, with the patient in the treatment position, and this information was registered with her original kV-CT scan for the purposes of structure delineation, dose calculation, and treatment planning. To validate the feasibility of the MV-FBCT-generated treatment plan, an electron density CT phantom (model 465, Gammex Inc., Middleton, WI) was scanned using MV-FBCT to obtain CT number to density table. Additionally, both a “cheese” phantom (which came with the tomotherapy treatment machine) with 2 inserted ion chambers and a generic phantom called Quasar phantom (Modus Medical Devices Inc., London, ON, Canada) with one inserted chamber were used to confirm dosimetric accuracy. The MV-FBCT could be used to clearly visualize anatomy in the region of the dental artifact and provide sufficient soft-tissue contrast to assist in the delineation of normal tissue structures and fat planes. With the elimination of the dental artifact, the MV-FBCT images allowed more accurate dose calculation by the tomotherapy system. It was confirmed that the phantom material density was determined correctly by the tomotherapy MV-FBCT number to density table. The ion chamber measurements agreed with the calculations from the MV-FBCT generated phantom plan within 2%. MV-FBCT may be useful in radiation treatment planning for nasopharyngeal cancer patients in the setting of extensive dental artifacts.
ISSN:0958-3947
1873-4022
DOI:10.1016/j.meddos.2009.03.006