Secondary Neutron Doses for Several Beam Configurations for Proton Therapy
Purpose To compare possible neutron doses produced in scanning and scattering modes, with the latter assessed using a newly built passive-scattering proton beam line. Methods and Materials A 40 × 30.5 × 30-cm water phantom was irradiated with 230-MeV proton beams using a gantry angle of 270°, a 10-c...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 2009-05, Vol.74 (1), p.260-265 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose To compare possible neutron doses produced in scanning and scattering modes, with the latter assessed using a newly built passive-scattering proton beam line. Methods and Materials A 40 × 30.5 × 30-cm water phantom was irradiated with 230-MeV proton beams using a gantry angle of 270°, a 10-cm-diameter snout, and a brass aperture with a diameter of 7 cm and a thickness of 6.5 cm. The secondary neutron doses during irradiation were measured at various points using CR-39 detectors, and these measurements were cross-checked using a neutron survey meter with a 22-cm range and a 5-cm spread-out Bragg peak. Results The maximum doses due to secondary neutrons produced by a scattering beam-delivery system were on the order of 0.152 mSv/Gy and 1.17 mSv/Gy at 50 cm from the beam isocenter in the longitudinal (0°) and perpendicular (90°) directions, respectively. The neutron dose equivalent to the proton absorbed dose, measured from 10 cm to 100 cm from the isocenter, ranged from 0.071 mSv/Gy to 1.96 mSv/Gy in the direction of the beam line ( i.e., φ = 0°). The largest neutron dose, of 3.88 mSv/Gy, was observed at 135° and 25 cm from the isocenter. Conclusions Although the secondary neutron doses in proton therapy were higher when a scattering mode rather than a scanning mode was used, they did not exceed the scattered photon dose in typical photon treatments. |
---|---|
ISSN: | 0360-3016 1879-355X |
DOI: | 10.1016/j.ijrobp.2008.10.090 |