Results of recent tests of NEC AMS systems

Results of the performance tests of the latest NEC 5 MV Accelerator Mass Spectrometry (AMS) systems at the Japan Nuclear Cycle Development Institute Tono Geoscience Center (JNC) in Gifu, Japan and at the National Institute for Environmental Studies (NIES) in Tsukuba, Japan are reported. Carbon datin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Sundquist, M. L., Loger, R. L., Daniel, R. E., Kitchen, R. L., Kolonko, J. J., Klody, G. M.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Results of the performance tests of the latest NEC 5 MV Accelerator Mass Spectrometry (AMS) systems at the Japan Nuclear Cycle Development Institute Tono Geoscience Center (JNC) in Gifu, Japan and at the National Institute for Environmental Studies (NIES) in Tsukuba, Japan are reported. Carbon dating precision on the order of 0.1 to 0.2% has been shown for solid and gas samples on both systems, using both the 40 sample MC-SNICS source with the NEC sequential injection system and the 12 sample MGF-SNICS gas source with the NEC simultaneous injection system. These results were obtained with analyzed beam currents up to 25 particle {mu}A (p{mu}A) {sup 12}C.
ISSN:0094-243X
1551-7616
DOI:10.1063/1.59316