A Technique for Verification of Isocenter Position in Tangential Field Breast Irradiation

Abstract Treatment verification and reproducibility of the breast treatment portals play a very important role in breast radiotherapy. We propose a simple technique to verify the planned isocenter position during treatment using an electronic portal imaging device. Ten patients were recruited in thi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2009, Vol.34 (1), p.16-19
Hauptverfasser: Prabhakar, Ramachandran, Ph.D, Pande, Manish, M.D, Harsh, Kumar, M.D, Julka, Pramod K., M.D, Ganesh, Tharmar, Ph.D, Rath, Goura K., M.D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Treatment verification and reproducibility of the breast treatment portals play a very important role in breast radiotherapy. We propose a simple technique to verify the planned isocenter position during treatment using an electronic portal imaging device. Ten patients were recruited in this study and (CT) computed tomography-based planning was performed with a conventional tangential field technique. For verification purposes, in addition to the standard medial (F1) and lateral (F2) tangential fields, a field (F3) perpendicular to the medial field was used for verification of the treatment portals. Lead markers were placed along the central axis of the 2 defined fields (F1 and F3) and the separation between the markers was measured on the portal images and verified with the marker separation on the digitally reconstructed radiographs (DRRs). Any deviation will identify the shift in the planned isocenter position during treatment. The average deviation observed between the markers measured from the DRR and portal image was 1.6 and 2.1 mm, with a standard deviation of 0.4 and 0.9 mm for fields F1 and F3, respectively. The maximum deviation observed was 3.0 mm for field F3. This technique will be very useful in patient setup for tangential breast radiotherapy.
ISSN:0958-3947
1873-4022
DOI:10.1016/j.meddos.2007.07.002