Intraaortic stabilization of the AneuRx stent-graft system: A useful adjunct to minimize device migration

The deployment of a Medtronic AneuRx stent-graft system for endovascular abdominal aortic aneurysm repair requires a series of precise maneuvers that include positioning the primary delivery catheter in the infrarenal aorta, retracting the graft-covering sheath, and withdrawing the stainless steel r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular and interventional radiology 2003-01, Vol.26 (1), p.73-75
Hauptverfasser: LIN, Peter H, MARTIN, Lou, MATSUURA, John H, WELLONS, Eric, BUSH, Ruth L, TERRAMANI, Thomas T, CHANGYI CHEN, LUMSDEN, Alan B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The deployment of a Medtronic AneuRx stent-graft system for endovascular abdominal aortic aneurysm repair requires a series of precise maneuvers that include positioning the primary delivery catheter in the infrarenal aorta, retracting the graft-covering sheath, and withdrawing the stainless steel runners. The last step allows the stent-graft to fully expand and attach to the non-aneurysmal aorta and iliac arteries. Such maneuvers may cause the stent-graft to move caudally if the device is placed in a severely angulated aortic neck. We describe a simple, coaxial, stabilization technique utilizing the contralateral introducer sheath which minimizes potential caudal migration of the stent-graft in angulated aortic necks during runner withdrawal.