Transvenous transjugular renal core biopsy with a redesigned biopsy set including a blunt-tipped needle
A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological dia...
Gespeichert in:
Veröffentlicht in: | Cardiovascular and interventional radiology 2002-03, Vol.25 (2), p.155-157 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological diagnosis. Immediate post-biopsy nonenhanced CT was performed to evaluate for bleeding at the biopsy site. All patients were observed for 2 hr after the procedure. No clinically significant immediate or late complication was noted. |
---|---|