The Impact of Advanced Technologies on Treatment Deviations in Radiation Treatment Delivery

Purpose To assess the impact of new technologies on deviation rates in radiation therapy (RT). Methods and Materials Treatment delivery deviations in RT were prospectively monitored during a time of technology upgrade. In January 2003, our department had three accelerators, none with “modern” techno...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of radiation oncology, biology, physics biology, physics, 2007-12, Vol.69 (5), p.1579-1586
Hauptverfasser: Marks, Lawrence B., M.D, Light, Kim L., C.M.D., R.T.T, Hubbs, Jessica L., M.S, Georgas, Debra L., R.T.T, Jones, Ellen L., M.D., Ph.D, Wright, Melanie C., Ph.D, Willett, Christopher G., M.D, Yin, Fang Fang, Ph.D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To assess the impact of new technologies on deviation rates in radiation therapy (RT). Methods and Materials Treatment delivery deviations in RT were prospectively monitored during a time of technology upgrade. In January 2003, our department had three accelerators, none with “modern” technologies (e.g., without multileaf collimators [MLC]). In 2003 to 2004, we upgraded to five new accelerators, four with MLC, and associated advanced capabilities. The deviation rates among patients treated on “high-technology” versus “low-technology” machines (defined as those with vs. without MLC) were compared over time using the two-tailed Fisher's exact test. Results In 2003, there was no significant difference between the deviation rate in the “high-technology” versus “low-technology” groups (0.16% vs. 0.11%, p = 0.45). In 2005 to 2006, the deviation rate for the “high-technology” groups was lower than the “low-technology” (0.083% vs. 0.21%, p = 0.009). This difference was caused by a decline in deviations on the “high-technology” machines over time ( p = 0.053), as well as an unexpected trend toward an increase in deviations over time on the “low-technology” machines ( p = 0.15). Conclusions Advances in RT delivery systems appear to reduce the rate of treatment deviations. Deviation rates on “high-technology” machines with MLC decline over time, suggesting a learning curve after the introduction of new technologies. Associated with the adoption of “high-technology” was an unexpected increase in the deviation rate with “low-technology” approaches, which may reflect an over-reliance on tools inherent to “high-technology” machines. With the introduction of new technologies, continued diligence is needed to ensure that staff remain proficient with “low-technology” approaches.
ISSN:0360-3016
1879-355X
DOI:10.1016/j.ijrobp.2007.08.017