Verification of In-Treatment Tumor Position Using Kilovoltage Cone-Beam Computed Tomography: A Preliminary Study
Three-dimensional tumor position during rotational dose delivery was evaluated by acquiring in-treatment kilovoltage (kV) cone-beam CT (CBCT) to ensure treatment quality. The CBCT projection data of a phantom were acquired during rotational megavoltage (MV) dose delivery up to 6 Gy to evaluate image...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 2007-11, Vol.69 (4), p.970-973 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Three-dimensional tumor position during rotational dose delivery was evaluated by acquiring in-treatment kilovoltage (kV) cone-beam CT (CBCT) to ensure treatment quality. The CBCT projection data of a phantom were acquired during rotational megavoltage (MV) dose delivery up to 6 Gy to evaluate image quality under MV beam irradiation. A lung tumor patient was treated with a total dose of 48 Gy in four fractions, each fraction including seven coplanar and noncoplanar beams, as well as a full-angle rotational beam. Tumor registration was performed between a planning CT image and a CBCT image immediately after patient setup. The patient couch was adjusted according to the registration results, and then the registration was repeated three times: immediately before treatment, during treatment, and immediately after treatment. The phantom image quality of the kV CBCT was not visually degraded up to the rotational MV dose of 6 Gy. Tumor position during rotational dose delivery was verified for the first time using kV CBCT. |
---|---|
ISSN: | 0360-3016 1879-355X |
DOI: | 10.1016/j.ijrobp.2007.08.026 |