Visible bremsstrahlung tomographic diagnostic for the pulsed high density field-reversed configuration experiment

A diagnostic suite for the source section of the pulsed high density field-reversed configuration (FRC) experiment has been constructed to investigate the equilibrium and stability of FRC plasmas. In particular, a visible bremsstrahlung tomographic system has been designed and implemented. Three typ...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2006-10, Vol.77 (10), p.10F319-10F319-5
Hauptverfasser: Gota, H., Andreason, S. P., Votroubek, G. R., Pihl, C. J., Slough, J. T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A diagnostic suite for the source section of the pulsed high density field-reversed configuration (FRC) experiment has been constructed to investigate the equilibrium and stability of FRC plasmas. In particular, a visible bremsstrahlung tomographic system has been designed and implemented. Three types of tomographic analyses for FRCs are performed: a Fourier fit method (Cormack-Granetz), a maximum entropy method, and a minimum Fisher method utilizing code developed for the TCV tokamak experiment in Switzerland [ Anton , Plasma Phys. Controled Fusion 38 , 1849 ( 1996 ) ]. Results from the different methods and end-on imaging from the fast-framing camera are compared showing relative agreement of FRC internal structures between all measurements.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.2220018