High-energy point-projection radiography of a driven, shielded Hohlraum
Point projection backlighting is a high magnification, high resolution x-ray backlighting configuration that provides detailed images of large objects at a reduced energy budget compared to area backlighting. However, point projection backlighting has the drawback that the camera aperture is large a...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2006-10, Vol.77 (10), p.10E324-10E324-3 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Point projection backlighting is a high magnification, high resolution x-ray backlighting configuration that provides detailed images of large objects at a reduced energy budget compared to area backlighting. However, point projection backlighting has the drawback that the camera aperture is large and open to more sources of noise. This issue has made it difficult to image driven targets without shielding. We present
9
keV
x-ray images of the rear end of a driven
Hohlraum
that employs shielding. We also present data of a
10.3
keV
x-ray image of an undriven
Hohlraum
. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.2351881 |