Application of Resistive Bolometer to Measurement of X-Ray Radiation from Laser Plasmas

A metal resistor thin-film bolometer was fabricated using the micro lithographic method and then applied to measurement of x-ray emission from gas-puff targets in the experiments at 1-kJ PALS laser in Prague. The detection head with the bolometer is equipped with a precise system for its calibration...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Ry, L, MacHa, P, Bartnik, A, Juha, L, Kaczmarczyk, J, Krasa, J, Miko, J, Bittner, M, Pfeifer, M, Ullschmied, J
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A metal resistor thin-film bolometer was fabricated using the micro lithographic method and then applied to measurement of x-ray emission from gas-puff targets in the experiments at 1-kJ PALS laser in Prague. The detection head with the bolometer is equipped with a precise system for its calibration. The bolometer is of 4-mum Au capturing layer and 100 % efficiency up to 5 keV. The lowest measurable flux is 5X10-9 J/cm2.