Application of Resistive Bolometer to Measurement of X-Ray Radiation from Laser Plasmas
A metal resistor thin-film bolometer was fabricated using the micro lithographic method and then applied to measurement of x-ray emission from gas-puff targets in the experiments at 1-kJ PALS laser in Prague. The detection head with the bolometer is equipped with a precise system for its calibration...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A metal resistor thin-film bolometer was fabricated using the micro lithographic method and then applied to measurement of x-ray emission from gas-puff targets in the experiments at 1-kJ PALS laser in Prague. The detection head with the bolometer is equipped with a precise system for its calibration. The bolometer is of 4-mum Au capturing layer and 100 % efficiency up to 5 keV. The lowest measurable flux is 5X10-9 J/cm2. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.2168840 |