Comparison of small photon beams measured using radiochromic and silver-halide films in solid water phantoms

In this study, we compared the dosimetric properties of four of the most commonly used films for megavoltage photon-beam dosimetry when irradiated under identical conditions by small multileaf-collimator (MLC) defined beamlets. Two silver-halide films (SHFs), Kodak XV2 and EDR2, and two radiochromic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2004-10, Vol.31 (10), p.2730-2737
Hauptverfasser: Zeidan, Omar A., Li, Jonathan G., Low, Daniel A., Dempsey, James F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, we compared the dosimetric properties of four of the most commonly used films for megavoltage photon-beam dosimetry when irradiated under identical conditions by small multileaf-collimator (MLC) defined beamlets. Two silver-halide films (SHFs), Kodak XV2 and EDR2, and two radiochromic films (RCFs), Gafchromic HS and MD55-2, were irradiated by MLC-defined 1 × 1 cm 2 beamlets from a Varian 2100 C ∕ D linac equipped with a 120-leaf MLC. The beamlets were delivered with the accelerator gantry set laterally ( 90 ° rotation) upon a solid-water compression film phantom at 100 cm source-to-surface distance which was positioned with the films parallel to the beam axis. Beamlets were delivered at central axis, 5.0 cm , and 10.5 cm off-axis for both leaf-end and leaf-side defined beamlets. The film dosimetry was performed using a quantitative optical density (OD) imaging system that was validated in a previous study. No significant differences between SHF and RCF measurements were observed in percentage depth doses, horizontal depth profiles, or two-dimension spatial isodose distributions in both the central axis and off-axis measurements. We found that regardless of the type of film used, RCF or SHF, a consistent data set for small beam dose modeling was generated. Previous validation studies based on the use of RCF and OD imaging system would indicate that all film produce an accurate result for small beam characterization.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.1788931