Measurement of Mass-Independent Perturbations of Cadmium in Nuclear Debris Samples
The isotopic composition of cadmium in nuclear debris was measured by multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS). Mass-independent perturbations due to fission product decay on the isotopes of Cd were observed. Relative fission yields for masses 111, 112, 114 and 116 we...
Gespeichert in:
Veröffentlicht in: | Geostandards and geoanalytical research 2022-12, Vol.47 (2) |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The isotopic composition of cadmium in nuclear debris was measured by multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS). Mass-independent perturbations due to fission product decay on the isotopes of Cd were observed. Relative fission yields for masses 111, 112, 114 and 116 were determined in each sample of nuclear debris and compared with reported values found in irradiated plutonium and uranium fuels. Mass spectrometry measurements of the valley fission product endpoints are extremely challenging due to the low cumulative fission yields and small amounts of each isotope that are produced. To overcome these challenges, a new purification method for Cd was developed and validated by examination of a number of geological reference materials. |
---|---|
ISSN: | 1639-4488 1751-908X |