Calibration of high voltages at the ppm level by the difference of $$^{83{\mathrm{m}}}$$Kr conversion electron lines at the KATRIN experiment

The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at - 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise high-voltage dividers were developed and built in cooperation with the German national metrology institute Phys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The European physical journal. C, Particles and fields Particles and fields, 2018-05, Vol.78 (5), Article 368
Hauptverfasser: Arenz, M, Baek, W.-J, Beck, M, Beglarian, A, Behrens, J, Bergmann, T, Berlev, A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at - 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise high-voltage dividers were developed and built in cooperation with the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB). Until now, regular absolute calibration of the voltage dividers required bringing the equipment to the specialised metrology laboratory. Here we present a new method based on measuring the energy difference of two [Formula omitted]Kr conversion electron lines with the KATRIN setup, which was demonstrated during KATRIN's commissioning measurements in July 2017. The measured scale factor [Formula omitted] of the high-voltage divider K35 is in agreement with the last PTB calibration 4 years ago. This result demonstrates the utility of the calibration method, as well as the long-term stability of the voltage divider.
ISSN:1434-6044
1434-6052
DOI:10.1140/epjc/s10052-018-5832-y